Smart-sensor toxic gas transmitter with true gas Auto-Test
Fixed toxic gas transmitter that takes 60 plug-in ATI smart sensors. It is available for 12–30 V DC, 115 V AC or 230 V AC with alarm relays, HART or Modbus RTU, heated, remote or duct-mount sensors, and Auto-Test gas generators that bump test the sensor automatically.
Auto-Test generates real test gas at the sensor every 24 hours, or on a schedule you set. If the sensor fails to respond, it raises a fault alarm, and you no longer need manual bump tests.
Smart sensors store their own zero and span history. Calibrate them in the shop, then plug them in with no transmitter adjustment.
One transmitter takes 60 sensor modules, the same family used in the PortaSens III portable.
Choose 12–30 V DC loop power, or 115/230 V AC with three alarm relays, plus HART or Modbus RTU.
Heated, 6 ft extension, remote junction box (up to 100 ft), and duct-mount sensor options.
Ordered separately. The configurator adds the interconnect cable or duct adapter that your sensor holder needs.
Calibration adapter
Flow cell assembly
Flow cell with 03-0460 sensor cap (for adding a flow cell to an existing F12/D)
Sensing module keeper for 4 sensors
Duct sensor adapter, 1½ in. MNPT
4-conductor interconnect cable, specify length (max. 100 ft)
6-conductor interconnect cable, specify length (max. 100 ft)
Panel mount bracket kit
Typical facilities and process areas where this product protects people and assets.
Key parameters below. For full datasheets, ratings, and ordering options, contact GDS.
| Manufacturer | Analytical Technology (ATI) |
| Category | Toxic Gas Monitors |
| Sensor type | Electrochemical smart sensor (60 modules available) |
| Range | User adjustable within the limits of the selected sensor |
| Accuracy | Generally ±10% of value (limited by calibration gas accuracy) |
| Repeatability / linearity | ±1% / ±0.5% (electronic) |
| Zero drift | Less than 2% of full scale per month |
| Span drift | Generally less than 3% per month (environment dependent) |
| Analog output | 4–20 mA; 700 Ω max. @ 24 V DC, 100 Ω max. @ 12 V DC |
| Serial interface | Optional HART on the 4–20 mA loop, or Modbus RTU over RS-232/485 |
| DC power | 2-wire loop, 25 mA max. @ 12–30 V DC; +50 mA @ 24 V DC with Modbus RS-485; +93 mA @ 24 V DC with heated sensor |
| AC power | 120 V AC or 220 V AC, 50/60 Hz, 7 W max. |
| Display | 96 × 32 dot-matrix graphic LCD, backlit, transflective |
| Controls | Four dome-type push buttons; remote alarm reset input (with alarm relays) |
| Enclosure | IP65 polycarbonate with stainless steel hardware |
| Mounting | Wall or pipe mount bracket (U-bolts for 1.5 or 2 in. pipe); optional panel mount kit |
| Operating environment | −22 to 140 °F (−30 to 60 °C), O₂ sensor min. 14 °F (−10 °C); 10–95% RH non-condensing |
| Weight | 1.5 lb (0.68 kg) |
| Approvals | CE, RoHS |